Company logo
BASH DVDSearchContentsHome
 
 
AVANCE III with IPSO
AQS/3
The main difference of AVANCE III from earlier series of spectrometer consoles is the IPSO (Intelligent Pulse Sequence Organizer) and the more extensive integration of different spectrometer functions into one device. The IPSO compatible SGU/2 is the successor of the former SGU with a much faster and wider digital interface for setting all relevant pulse program parameters, such as frequency, phase and amplitude. A package of frequency, phase and amplitude words, together with some relevant real-time pulses can be set synchronously every 25 ns via the 80 MHz and 48-bit wide LVDS link. The setting via the synchronous and differential LVDS link guarantees an excellent isolation between digital computer electronics and purest RF signals.
The SGU/2 incorporates many more features, such as a pre-viewer for all hardware relevant pulses. All pre- and post-delayed pulses such as transmitter, preamplifier and receiver gating are generated automatically and must be initialized only once. Therefore, one can fully focus on the NMR sequence and only has to deal with 'ideal' pulses.
The AQS/3 chassis is the consecutive development of the well proven AQS/2 chassis and can be equipped with the IPSO AQS Unit. The extended IPSO 19" Unit represents a standalone unit and doesn't fit into the AQS/3 chassis. The AQS IPSO is used with the NanoBay and MicroBay and the extended IPSO 19" Unit with the OneBay and TwoBay.
Viewing the AQS/3 chassis from the front, on the left is a proprietary IPSO backplane incorporated. On the right, the extended AQS/3 user bus is able to supply a lot of different AQS functional units (signal generators, receivers, preamplifiers, amplifiers, etc.).
The IPSO 19" Unit is a standalone unit that can be placed independently in the console rack. The unit has an internal power supply, supports up to a maximum of 8 channels, contains 2 additional PCI slots, 4 RS232 ports and 2 USB ports.
BSMS/2 with ELCB
The BSMS/2 with ELCB units contain a set of higher integrated boards (ELCB and SCB20) which have replaced the former controller, lock and shim boards (CPU/3, LCB, all variants of SCB7/13). These boards provide better performance, higher resolution and increased stability. The shim boards have been unified - a standard SCB20 now provides the necessary precision for many variants of connected shim systems.
Probes
General information common to most Probes can be found here. See the appropriate NMR probe documentation for details.
HPPR/2 preamplifier module
The HPPR/2 amplifies, filters and routes the NMR response signals from the probe to the RX22 receiver. It switches the RF transmitter output to the probe.
Magnet
General handling and maintenance information for all magnets can be found here. Specific documents are supplied with each magnet.
DNP-NMR (optional)
DNP NMR allows the characterization, at the molecular level, of hybrid organic silica material. These materials are key compounds for applications in catalysis, drug delivery, separation and purification devices. The DNP Control System Console controls and supervises the unique high power microwave source and its infrastructure.
GRASP (optional)
The new GREAT for AVANCE III no longer applies preemphasis to the gradients on its own. Instead, the preemphasis is digitally computed and applied to the digital gradient data by an external unit, the DPP (Digital Preemphasis Processor). The preemphasis is now calculated by a DSP and works on the digital input of the GREAT.
Automation (optional)
Bruker Automation allows the user to easily and effectively perform automatic experiments.
Further information
AVANCE III with IPSO - NanoBay Cabinet
AVANCE III with IPSO - MicroBay Cabinet
AVANCE III with IPSO - OneBay Cabinet
AVANCE III with IPSO - TwoBay Cabinet